HP Data Explorer 4 Series User Manual

Page 332

Advertising
background image

Chapter 7 Data Explorer Examples

7-16

Applied Biosystems

7

Calibrating

To calibrate the deisotoped spectrum:

1.

From the Peaks menu, select Peak Label, and select
the Mass Label Type (peak apex or peak centroid) to
use for calibration. Click OK.

2.

From the Process menu, select Mass Calibration and
then select Manual Calibration.

The Manual Calibration dialog box is displayed
(Figure 7-16).

Figure 7-16 Manual Calibration Dialog Box

3.

Click

, then select the VOYAGER.REF calibration

reference file.

4.

Enter Reference Matching and Calibration Criteria.

NOTE: For descriptions of calibration parameters, see
Section 5.3, Manual Calibration.