Chapter 9 Troubleshooting

9-12

Applied Biosystems

9

Table 9-7 Calibration Troubleshooting—Mariner Only

Symptom

Possible Cause

Action

Mass Calibration
commands are dimmed

Chromatogram window is
selected

Select Spectrum window.

Apply Calibration
command is dimmed
when calibrating
MS Method data

For MS Method data,
calibration is valid for an
individual spectrum. You
cannot apply the
calibration from one
spectrum to the entire
data file.

No action. Normal
occurrence.

Error displayed when you
import a calibration

.CAL file corrupted

Create new .CAL file. See
“Exporting .BIC, .MSM,
and .CAL files” on
page 1-36.

Importing a .CAL file
generated from a Voyager
data file

Import a Mariner .CAL.