Peak Detection and Labeling Troubleshooting

Data Explorer

Software User’s Guide

9-21

9

Spectrum peaks labeled
with incorrect charge state
when charge state labels
are selected

(continued)

Charge state
determination parameters
are set such that peaks
are determined to have no
charge

Adjust parameters. See:

“Peak Processing
parameters
(spectrum data only)”
on page 3-26

Section 3.2.5,
Charge State
Determination and
Examples

Noise between isotope
peaks is detected

Increase %Base Peak
Intensity or %Max Peak
Area to eliminate noise.
See “Basic Settings
(spectrum data only)” on
page 3-22.

Apply noise filter. See
Section 5.7, Noise
Filtering/Smoothing.

Monoisotopic peak not
labeled correctly

Peak detection charge
state parameters are not
set correctly, the software
is identifying the tallest
peak in a cluster as the
monoisotopic peak

Adjust parameters to
correctly identify the
monoisotopic peak. See
“Peak Processing
parameters (spectrum
data only)” on page 3-26.

Table 9-12 Charge State and Isotope Determination Troubleshooting—Mariner

Only (Continued)

Symptom

Possible Cause

Action