HP Data Explorer 4 Series User Manual

Page 210

Advertising
background image

Chapter 5 Examining Spectrum Data

5-6

Applied Biosystems

5

The manual calibration feature provides two modes for peak
matching:

Automatic—The software automatically compares
reference masses to observed masses, and lists peaks
that are within the specified peak matching criteria.

Manual—You manually select an observed mass, then
select the reference mass for comparison.

Hint: It is often useful to automatically perform a match and
fit first, then manually adjust the fit as needed.

Calibration

references files

(.REF) provided

Default calibration reference files are provided with the
software. You can use these files as a starting point and add
delete reference masses as needed.

For information, see Section 5.3.3, Creating or Modifying a
Calibration Reference File (.REF)
.

When to use

manual

calibration

Use manual calibration when you:

Calibrate Voyager data

Calibrate Mariner data and any of the following occur:

You have one or only a few spectra to calibrate

You do not know in advance which reference
masses to use

You do not know in advance if the quality of the
reference mass signals is acceptable

You want to fine-tune the results of an automatic
calibration

For information on automatic calibration, see Section 5.4,
Automatic Calibration.

Accurate mass

measurements

Before performing an internal calibration, refer to the:

Mariner Workstation User’s Guide

Voyager Biospectrometry Workstation User’s Guide

Advertising