Temperature calibration – Garland MP-GD-10-S User Manual

Page 33

Advertising
background image

Part # MPSM07 (03/24/08)

Page 33

TEMPERATURE CALIBRATION

1 . Press and hold “PROGRAM” Button for three (3) seconds .

When prompted, enter the Diagnostics Code 3-4-2-4,

Press “Y” . The Moisture Plus part number will display briefly

followed by the Diagnostic Tests – Keypad Test .

2 . Press the “N” Button to scroll to Diagnostic Tests – Temp

Cal . Press “Y” .

3 After selecting Temperature Calibration and Pressing

“ENTER” . The display will show the Actual Temperature

inside the oven . Verify the cavity temperature using a

pyrometer .

4 . Press and hold the “PHANTOM” button for three, (3),

seconds . The display will show the Actual Temperature

on the first line and Set Temperature on the second line .

If the Actual Temperature does not correspond to the

pryometer, enter the value on the pyromter using buttons

1 – 0 . The value will be input as Set Temp .

5 . Press “Y” to accept the changes . The actual Temperature

will change to equal the value input for Set Temperature .

Press Cancel to exit Temperature Calibration .

NOTE: After the desired Tests are completer, press “PROGRAM”

to exit Diagnostics Mode .