Chapter 1 Introduction

Series 220/221 Instruction Manual

1-6

IM-22

As shown below, Innova-Flo

Vortex Flow Meters exhibit a constant

Strouhal number across a large range of Reynolds numbers, indicating a
consistent linear output over a wide range of flows and fluid types. Be-
low this linear range, the intelligent electronics in Innova-Flo automati-
cally corrects for the variation in the Strouhal number. Innova-Flo’s
smart electronics correct for this non-linearity by calculating the Rey-
nolds number based on constant values of the fluid’s density and viscos-
ity stored in the instrument’s memory. Innova-Flo Vortex Flow Meters
automatically correct down to a Reynolds number of 5,000.



0.3

0.2

0.1

0.0

3

4

10

Linear range

Reynolds Number, Re

Strouhal Number, St

10

5

10

10

8

10

6

10

7

5000

Corrected range

Figure 1-3. Reynolds Number Range for the Innova-Flo Meter