Chapter 4 HART Communications
Series 220/221 Instruction Manual
Sequence Description
Access
Notes
1,4,3,1,3,1
4 mA
View
Loop test, fix analog output at 4 mA
1,4,3,1,3,2
20 mA
View
Loop test, fix analog output at 20 mA
1,4,3,1,3,3 Other
Edit
Loop test, fix analog output at mA value entered
1,4,3,1,3,4 End
Exit loop test
1,4,3,1,4 D/A
trim
Edit
Calibrate electronics 4mA and 20mAvalues
1,4,3,1,5 Scaled
D/A
trim N/A Not
used
1,4,3,2,1 Poll
addr
Edit Poll
address
1,4,3,2,2
Num req. preams
View
Number of required preambles
1,4,3,2,3
Burst mode
N/A
Not used
1,4,3,2,4 Burst
option N/A
Not
used
1,4,4,1 Distributor
N/A
Not
used
1,4,4,2 Model
N/A
Not
used
1,4,4,3 Dev
id
View
Device identification
1,4,4,4 Tag
Edit
Tag
1,4,4,5 Date
Edit
Date
1,4,4,6 Write
Protect View Write
protect
1,4,4,7 Descriptor
Edit Vortex
flowmeter
1,4,4,8 Message
Edit
32
character alphanumeric message
1,4,4,9
PV snsr s/n
View
Primary variable sensor serial number
1,4,4,menu
Final assy #
Edit
Final assembly number
1,4,4,menu,1 Universal
Rev
View Universal
revision
1,4,4,menu,2
Fld dev Rev
View
Field device revision
1,4,4,menu,3
Software Rev
View
Software revision
1,5 Review N/A
Not
used
2
PV
View
Primary variable value
3
PV AO
View
Analog output, mA
4,1
PV LRV
Edit
Primary variable lower range value
4,2
PV URV
Edit
Primary variable upper range value
5,1
PV LRV
Edit
Primary variable lower range value
5,2
PV URV
Edit
Primary variable upper range value
4-4
IM-22