Chapter 3 Operation

Series 220/221 Instruction Manual

Diagnostics Menu


Password

ENTER

Use
keys to access menus

Diagnostics

Menu

Sim Vor Freq

xxx

ENTER

Run Mode

Highest Velocity

xxx

Simulate Vortex
Frequency (Hz)

Highest Recorded
Velocity (ft/sec)


Use the Diagnostics Menu to simulate flow and review the highest recorded veloc-
ity in ft/sec.

The simulated vortex frequency is used for testing the meter to verify that the
programming is correct. Enter any value for the sensor input in Hz. The meter
will calculate a flow rate based on the corresponding value and update the analog
output and totalizer pulse output. Note: when your diagnostic work is complete,
make sure to return the simulated frequency to zero to allow the electronics to
use the actual value.