Fast key sequence – Sierra INNOVA-FLO 220 User Manual
Page 50
Series 220/221 Instruction Manual
Chapter 4 HART Communications
Fast Key Sequence
Use password 16363.
Sequence Description
Access
Notes
1,1,1
Snsr
View
Primary variable value
1,1,2
AI % Rnge
View
Analog output % range
1,1,3
AO1
View
Analog output, mA
1,2,1
Test Device
N/A
Not used
1,2,2,1
4 mA
View
Loop test, fix analog output at 4 mA
1,2,2,2
20 mA
View
Loop test, fix analog output at 20 mA
1,2,2,3
Other
Edit
Loop test, fix analog output at mA value entered
1,2,2,4
End
Exit loop test
1,2,3,1,1 4
mA
N/A Not used, apply values
1,2,3,1,2 20
mA
N/A Not used, apply values
1,2,3,1,3 Exit
Exit apply values
1,2,3,2,1 PV
LRV
Edit Primary variable lower range value
1,2,3,2,2 PV
URV
Edit Primary variable upper range value
1,2,3,2,3 PV
USL
View
Primary variable upper sensor limit
1,2,3,2,4 PV
LSL
View
Primary
variable lower sensor limit
1,2,4 D/A
Trim Edit
Calibrate electronics 4mA and 20mAvalues
1,3,1 Tag
Edit
Tag
1,3,2
PV unit
Edit
Primary variable units
1,3,3,1 PV
LRV
Edit
Primary variable lower range value
1,3,3,2 PV
URV
Edit
Primary variable upper range value
1,3,3,3 PV
LSL
View
Primary
variable upper sensor limit
1,3,3,4 PV
USL
View
Primary
variable lower sensor limit
1,3,4,1 Distributor
N/A
Not
used
1,3,4,2 Model
N/A
Not
used
1,3,4,3 Dev
id
View
Device identification
1,3,4,4 Tag
Edit
Tag
1,3,4,5 Date
Edit
Date
1,3,4,6 Write
Protect View Write
protect
1,3,4,7 Descriptor
Edit Vortex
flowmeter
1,3,4,8 Message
Edit
32
character alphanumeric message
1,3,4,9
PV snsr s/n
View
Primary variable sensor serial number
1,3,4,menu
Final assy #
Edit
Final assembly number
1,3,4,menu,1 Universal
Rev
View Universal
revision
1,3,4,menu,2
Fld dev Rev
View
Field device revision
1,3,4,menu,3
Software Rev
View
Software revision
1,3,5
PV Xfer fnctn
View
Linear
1,3,6
PV Damp
Edit
Primary variable damping (time constant) in seconds
1,4,1,1
PV
View
Primary variable value
1,4,1,2
PV Sensor Unit
Edit
Primary variable units
1,4,1,3 Sensor
Information
View
PV LSL, PV USL, PV Min span
1,4,2,1
Snsr Damp
Edit
Primary variable damping (time constant) in seconds
1,4,2,2,1 PV
LRV
Edit Primary variable low range value
1,4,2,2,2 PV
URV
Edit Primary variable upper range value
1,4,2,3,1 PV
LRV
Edit Primary variable low range value
1,4,2,3,2 PV
URV
Edit Primary variable upper range value
1,4,2,4 Xfer
Fnctn View
Linear
1,4,2,5
AI % rnge
View
Analog output % range
1,4,3,1,1 AO1
View
Analog output, mA
1,4,3,1,2
AO alarm typ
N/A
Not used
Continued on next page
IM-22
4-3