Chapter 2 Installation

Series 220/221 Instruction Manual

Installing Flow Meters with a Compression Connection*

Use the following formula to determine insertion length for flow meters
(NPT and flanged) with a compression process connection. The installation
procedure is given on the next page.

Insertion Length Formula

I = S – F – R – t

Where:

I = Insertion length.

S = Stem length – the distance from the center of the sensor head to the base of the

enclosure adapter (S = 29.47 inches for standard probes; S = 13.1 inches for
compact; S = 41.47 inches for 12-inch extended).

F = Distance from the raised face of the flange or top of NPT stem housing to the

outside of the pipe wall.

R = Pipe inside diameter

÷ 2 for pipes ten inches and smaller.

R = Five inches for pipe diameters larger than ten inches.

t = Thickness of the pipe wall. (Measure the disk cut-out from the tapping proce-

dure or check a piping handbook for thickness.)

Figure 2-6. Insertion Calculation (Compression Type)

Example:

To install a Series 221 meter with a standard probe (S = 29.47 inches) into a 14
inch schedule 40 pipe, the following measurements are taken:

F = 3

inches

R = 5

inches

t =

0.438

inches


The insertion length for this example is 21.03 inches. Insert the stem through the fit-
ting until an insertion length of 21.03 inches is measured with a ruler.

*All dimensions are in inches