Digital Routing and Clock
Generation

The digital routing circuitry has the following three main functions.

Manages the flow of data between the bus interface and the
acquisition/generation sub-systems (analog input, analog output,
digital I/O, and the counters). The digital routing circuitry uses FIFOs
(if present) in each sub-system to ensure efficient data movement.

Routes timing and control signals. The acquisition/generation
sub-systems use these signals to manage acquisitions and generations.
These signals can come from the following sources.

Your M Series device

Other devices in your system through RTSI

User input through the PFI terminals

User input through the PXI_STAR terminal

Routes and generates the main clock signals for the M Series device.

Clock Routing

Figure 10-1 shows the clock routing circuitry of an M Series device.

Figure 10-1. M Series Clock Routing Circuitry

RTSI <0..7>

Onboard

80 MHz

Oscillator

External

Reference

Clock

(To RTSI <0..7>
Output Selectors)

10 MHz RefClk

PLL

ч

4

ч

200

ч 8

PXI_CLK10

PXI_STAR

80 MHz Timebase

100 kHz Timebase

20 MHz Timebase