Chapter 4

Analog Input

NI 6236 User Manual

4-14

ni.com

Figure 4-9. Posttriggered Data Acquisition Example

Pretriggered data acquisition allows you to view data that is acquired before
the trigger of interest, in addition to data acquired after the trigger.
Figure 4-10 shows a typical pretriggered DAQ sequence. ai/StartTrigger
can be either a hardware or software signal. If ai/StartTrigger is set up to be
a software start trigger, an output pulse appears on the ai/StartTrigger line
when the acquisition begins. When the ai/StartTrigger pulse occurs, the
sample counter is loaded with the number of pretriggered samples, in this
example, four. The value decrements with each pulse on ai/SampleClock,
until the value reaches zero. The sample counter is then loaded with the
number of posttriggered samples, in this example, three.

Figure 4-10. Pretriggered Data Acquisition Example

If an ai/ReferenceTrigger pulse occurs before the specified number of
pretrigger samples are acquired, the trigger pulse is ignored. Otherwise,
when the ai/ReferenceTrigger pulse occurs, the sample counter value
decrements until the specified number of posttrigger samples have been
acquired.

1

3

0

4

2

ai/StartTrigger

ai/SampleClock

ai/ConvertClock

Sample Counter

n/a

0

1

2

3

1

0

2

2

2

ai/StartTrigger

ai/ReferenceTrigger

ai/SampleClock

ai/ConvertClock

Scan Counter