Figure 7-29. filter example, Prescaling, Figure 7-30. prescaling – National Instruments DAQ M Series User Manual

Page 95: Prescaling -33

Advertising
background image

Chapter 7

Counters

© National Instruments Corporation

7-33

NI 6236 User Manual

The filter setting for each input can be configured independently. On power
up, the filters are disabled. Figure 7-29 shows an example of a low to high
transition on an input that has its filter set to 125 ns (N = 5).

Figure 7-29. Filter Example

Enabling filters introduces jitter on the input signal. For the 125 ns and
6.425 µs filter settings, the jitter is up to 25 ns. On the 2.55 ms setting, the
jitter is up to 10.025 µs.

When a PFI input is routed directly to RTSI, or a RTSI input is routed
directly to PFI, the M Series device does not use the filtered version of the
input signal.

Refer to the KnowledgeBase document, Digital Filtering with M Series, for
more information about digital filters and counters. To access this
KnowledgeBase, go to

ni.com/info

and enter the info code

rddfms

.

Prescaling

Prescaling allows the counter to count a signal that is faster than the
maximum timebase of the counter. M Series devices offer 8X and 2X
prescaling on each counter (prescaling can be disabled). Each prescaler
consists of a small, simple counter that counts to eight (or two) and rolls
over. This counter can run faster than the larger counters, which simply
count the rollovers of this smaller counter. Thus, the prescaler acts as a
frequency divider on the Source and puts out a frequency that is one-eighth
(or one-half) of what it is accepting.

Figure 7-30. Prescaling

1

2

3

1

4

1

2

3

4

5

RTSI, PFI, or

PXI_STAR Terminal

Filter Clock

(40 MHz)

Filtered Input

Filtered input goes high
when terminal is sampled
high on five consecutive
filter clocks.

External Signal

Counter Value

Prescaler Rollover

(Used as Source

by Counter)

0

1

Advertising