Chapter 4 analog input, Figure 4-1. ni 6236 analog input circuitry, Analog input circuitry – National Instruments DAQ M Series User Manual

Page 25: Analog input circuitry -1, Of chapter 4, Analog input

Advertising
background image

© National Instruments Corporation

4-1

NI 6236 User Manual

4

Analog Input

Figure 4-1 shows the analog input circuitry of NI 6236 devices.

Figure 4-1. NI 6236 Analog Input Circuitry

Analog Input Circuitry

I/O Connector

You can connect analog input signals to the M Series device through the I/O
connector. The proper way to connect analog input signals depends on the
analog input ground-reference settings, described in the

Analog Input

Ground-Reference Settings

section. Also refer to Appendix A,

NI 6236

Device Information

, for device I/O connector pinouts.

MUX

Each M Series device has one analog-to-digital converter (ADC). The
multiplexers (MUX) route one AI channel at a time to the ADC through the
NI-PGIA.

Instrumentation Amplifier (NI-PGIA)

The NI programmable gain instrumentation amplifier (PGIA) is a
measurement and instrument class amplifier that minimizes settling times
for all input ranges. The NI-PGIA can amplify or attenuate an AI signal to
ensure that you use the maximum resolution of the ADC.

I/O Connector

AI <0..3>+

AI GND

NI-PGIA

Input Range

Selection

ADC

AI Data

Isolation

Barrier

Digital

Isolators

AI FIFO

AI <0..3>–

Mux

Advertising