Chapter 5

Analog Output

NI 6236 User Manual

5-6

ni.com

Figure 5-3. Analog Output Timing Options

NI 6236 devices feature the following AO (waveform generation) timing
signals.

AO Start Trigger Signal

AO Pause Trigger Signal

AO Sample Clock Signal

AO Sample Clock Timebase Signal

AO Start Trigger Signal

Use the AO Start Trigger (ao/StartTrigger) signal to initiate a waveform
generation. If you do not use triggers, you can begin a generation with a
software command.

Using a Digital Source

To use ao/StartTrigger, specify a source and an edge. The source can be one
of the following signals:

A pulse initiated by host software

Input PFI <0..5>

RTSI <0..7>

ai/ReferenceTrigger

ai/StartTrigger

PXI_STAR

PFI, RTSI

PXI_STAR

20 MHz Timebase

100 kHz Timebase

PXI_CLK10

Programmable

Clock

Divider

ao/Sample

Clock

Timebase

PFI, RTSI

PXI_STAR

Ctr

n Internal Output

ao/Sample

Clock