Timing signal routing, Figure 3-9. convert* signal routing, Timing signal routing -15 – National Instruments VXI-MIO Series User Manual

Page 44: Figure 3-9. convert* signal routing -15

Advertising
background image

Chapter 3 Hardware Overview

National Instruments Corporation

3-15

VXI-MIO Series User Manual

Timing Signal Routing

The DAQ-STC provides a very flexible interface for connecting timing

signals to other modules or external circuitry. Your VXI-MIO Series
module uses the VXIbus trigger for interconnecting timing signals
between modules and the Programmable Function Input (PFI) pins on
the I/O connector for connecting to external circuitry. These
connections are designed to enable the VXI-MIO Series module to both
control and be controlled by other modules and circuits.

There are a total of 13 timing signals internal to the DAQ-STC that can

be controlled by an external source. These timing signals can also be
controlled by signals generated internally to the DAQ-STC, and these
selections are fully software-configurable. For example, the signal
routing multiplexer for controlling the CONVERT* signal is shown in
Figure 3-9.

Figure 3-9. CONVERT* Signal Routing

VXI TTL Trigger <0..4>

VXI ECL Trigger <0..1>

PFI<0..9>

CONVERT*

Sample Interval Counter TC

GPCTR0_OUT