Appendix A Specifications for VXI-MIO-64E-1

National Instruments Corporation

A-3

VXI-MIO Series User Manual

Transfer Characteristics

Relative accuracy .............................

±

0.5 LSB typ dithered,

±

1.5 LSB

max undithered

DNL ..................................................

±

0.5 LSB typ,

±

1.0 LSB max

No missing codes .............................. 12 bits, guaranteed

Offset error

Pregain error after calibration .....

±

12

µ

V max

Pregain error before calibration...

±

2.5 mV max

Postgain error after calibration ....

±

0.5 mV max

Postgain error before calibration .

±

100 mV max

Gain error (relative to calibration reference)

After calibration (gain = 1) .........

±

0.02% of reading max

Before calibration .......................

±

2.5% of reading max

Gain

1 with gain error

adjusted to 0 at gain = 1 .......

±

0.02% of reading max

Amplifier Characteristics

Input impedance

Normal powered on..................... 100 G

in parallel with 100 pF

Powered off ................................ 1 k

min

Overload ..................................... 1 k

min

Input bias current ..............................

±

200 pA

Input offset current ............................

±

100 pA

CMRR, DC to 60 Hz

Gain = 0.5 ................................... 95 dB

Gain = 1 ...................................... 100 dB

Gain

2 ...................................... 106 dB

Advertising