Scanclk signal, Extstrobe* signal, Figure 4-12. typical pretriggered acquisition – National Instruments VXI-MIO Series User Manual

Page 72: Figure 4-13. scanclk signal timing, Scanclk signal -26 extstrobe* signal -26

Advertising
background image

Chapter 4 Signal Connections

VXI-MIO Series User Manual

4-26

National Instruments Corporation

Figure 4-12.

Typical Pretriggered Acquisition

SCANCLK Signal

SCANCLK is an output-only signal that generates a pulse with the

leading edge occurring approximately 50 to 100 ns after an A/D
conversion begins. The polarity of this output is software selectable but
is typically configured so that a low-to-high leading edge can clock
external analog input multiplexers indicating when the input signal has
been sampled and can be removed. This signal has a 400 to 500 ns pulse
width and is software-enabled. Figure 4-13

shows the timing for the

SCANCLK signal.

Figure 4-13.

SCANCLK Signal Timing

EXTSTROBE* Signal

EXTSTROBE* is an output-only signal that generates either a single

pulse or a sequence of eight pulses in the hardware-strobe mode. An
external device can use this signal to latch signals or to trigger events.
In the single-pulse mode, your software controls the EXTSTROBE*

Don't Care

0

1

2

3

1

0

2

2

2

TRIG1

TRIG2

STARTSCAN

CONVERT*

Scan Counter

t

w

t

w

= 400 to 500 ns

t

d

= 50 to 100 ns

t

d

CONVERT*

SCANCLK

Advertising