Startscan signal, Figure 4-19. startscan input signal timing, Startscan signal -30 – National Instruments VXI-MIO Series User Manual

Page 76: Figure 4-19. startscan input signal timing -30

Advertising
background image

Chapter 4 Signal Connections

VXI-MIO Series User Manual

4-30

National Instruments Corporation

STARTSCAN Signal

Any PFI pin can externally input the STARTSCAN signal, which is

available as an output on the PFI7/STARTSCAN pin.

Refer to Figures 4-11 and 4-12 for the relationship of STARTSCAN to

the data acquisition sequence.

As an input, the STARTSCAN signal is configured in the

edge-detection mode. You can select any PFI pin as the source for
STARTSCAN and configure the polarity selection for either rising or
falling edge. The selected edge of the STARTSCAN signal initiates a
scan. The sample interval counter starts if you select internally triggered
CONVERT*.

As an output, the STARTSCAN signal reflects the actual start pulse that

initiates a scan. This is true even if the starts are being externally
triggered by another PFI. You have two output options. The first is an
active high pulse with a pulse width of 50 to 100 ns, which indicates the
start of the scan. The second action is an active high pulse that
terminates at the start of the last conversion in the scan, which indicates
a scan in progress. STARTSCAN will be deasserted t

off

after the last

conversion in the scan is initiated. This signal is set to input (High-Z) at
startup.

Figures 4-19 and 4-20 show the input and output timing requirements

for the STARTSCAN signal.

Figure 4-19.

STARTSCAN Input Signal Timing

Rising-edge
polarity

Falling-edge
polarity

t

w

t

w

= 10 ns minimum

Advertising