Chapter 5

Programming

SCXI-1121 User Manual

5-14

www.natinst.com

5.

Write binary

101S 1100

to the HSCR.

6.

Write binary

101S 1110

to the HSCR.

7.

Write binary

101S 1111

to the HSCR.

In the preceding steps, S = 0 if you want the scanning to repeat when the
end of the list is reached. S = 1 if you want the circuitry to shut down after
a single scan.

When you are writing multiple entries to the same register, for example,
repetitive writes to the HSCR or several FIFO entries, it is important that
SS*13 or SS*14 go inactive (high) between each entry. Select another slot
or toggle the SLOT0SEL* line to temporarily deassert the appropriate SS*
line.

If consecutive scan list entries access an SCXI-1121, the module will reload
the MUXCOUNTER with the starting channel after each entry. Thus, two
entries with counts of two for one module will yield different behavior than
one entry with a count of four.

For multiple-chassis scanning, program each Slot 0 with dummy entries to
fill the sample counts when the data acquisition board is accessing other
chassis. Use Slot 13 as the dummy entry slot.

See

Example 3

at the end of this chapter.

4. Acquisition Enable, Triggering, and Servicing

At this point, you should now continue from where you left off in the

1. Data Acquisition Board Setup Programming

section of this chapter.

Perform the following steps given in your data acquisition board user
manual.

MIO Board User Manual

Enable the scanning data acquisition operation.

Apply a trigger.

Service the data acquisition operation.

Lab-PC User Manual, Lab-PC+ User Manual, and PC-LPM-16 User
Manual

Start and service the data acquisition operation.

Lab-LC User Manual

Program the sample-interval counter.

Service the data acquisition operation.