Chapter 2 Configuration and Installation

SCXI-1121 User Manual 2-42 www.natinst.com

Timing Requirements and Communication Protocol

Timing Signal

The data acquisition timing signal is SCANCLK.

SCANCLK is used to increment MUXCOUNTER on its rising edge.
Figure 2-15 shows the timing requirements of the SCANCLK signal. These
requirements will ensure that SCANCLK is properly transmitted over
TRIG0.

Figure 2-15. SCANCLK Timing Requirements

For output selection time specifications, refer to Appendix A,

Specifications

.

Communication Signals

This section describes the methods for communicating on the Serial
Peripheral Interface (SPI) bus and their timing requirements. The
communication signals are SERDATIN, DAQD*/A, SLOT0SEL*,
SERDATOUT, and SERCLK. Furthermore, SS* is produced by Slot 0
according to data acquisition board programming, and SS* timing
relationships will also be discussed. For information on the Slot 0
Slot-Select Register, consult Chapter 4,

Register Descriptions

.

The data acquisition board determines to which slot it will talk by writing
a slot-select number to Slot 0. In the case of an SCXI-1001 chassis, this
write also determines to which chassis the data acquisition board will talk.
Writing a slot-select number is also used in programming the Slot 0
hardscan circuitry. See Chapter 5,

Programming

, for information on

programming the Slot 0 hardscan circuitry.

The following sections detail the procedure for selecting a slot in a
particular chassis. Figure 2-16 illustrates the timing of this procedure with
the example case of selecting Slot 11 in Chassis 9. Notice that the
factory-default chassis address for the SCXI-1000 is address 0. For
information on changing the address of your chassis, consult the
SCXI-1000/1001 User Manual. An SCXI-1000 chassis will respond to any
chassis number.

T

low

T

high

SCANCLK

T

low

Time low before rising edge 400 nsec minimum

T

high

Time high before falling edge 250 nsec minimum