2.

Following the procedure given in the

Register Writes

section, write

00000000 00000000

to the Configuration Register of the

SCXI-1121 in Slot 2. This step resets the module, including the
clearing of the AB0EN bit (bit 0). Notice that a complete reset of this
module is not necessary, but is used for simplicity.

3.

Following the procedure given in the

Register Writes

section, write

100XXX00 00001111

to the Configuration Register of the

SCXI-1121 in Slot 4.

4.

Following the procedure given in the

Register Writes

section, write

010XXX01 00001111

to the Configuration Register of the

SCXI-1121 in Slot 1.

5.

Following the procedure given in the

Register Writes

section, write

010XXX11 00001111

to the Configuration Register of the

SCXI-1121 in Slot 3. Notice that after Channel 3, the SCXI-1121 will
wrap around to Channel 0.

6.

Follow the steps given in the section earlier in this chapter,

3. Programming the Slot 0 Hardscan Circuitry

, where step 3, Write the

Slot 0 scan list to the FIFO, consists of the following:

a.

Write

00000000 00000010

to the FIFO Register. This

corresponds to Slot 1 for three samples.

b.

Write

00000001 10000001

to the FIFO Register. This

corresponds to Slot 4 for two samples.

c.

Write

00000001 00000011

to the FIFO Register. This

corresponds to Slot 3 for four samples.

Make sure to toggle SLOT0SEL* or reselect the FIFO Register from
scratch between steps 6a, 6b, and 6c.

7.

Follow the procedure given in the

4. Acquisition Enable, Triggering,

and Servicing

section earlier in this chapter.

Example 3

You want to scan four channels on an SCXI-1121 in Slot 4 of Chassis 1,
then two channels of an SCXI-1121 in Slot 11 of Chassis 2, one channel of
an SCXI-1121 in Slot 3 in Chassis 3, and three channels of an SCXI-1121
in Slot 8 of Chassis 3.

Advertising