Table 2-16 – National Instruments SCXI-1121 User Manual

Page 57

Advertising
background image

Chapter 2

Configuration and Installation

© National Instruments Corporation

2-41

SCXI-1121 User Manual

The digital timing signals are pins 36 and 43.

Pin 36 is used as a clock by the SCXI-1121 to increment the
MUXCOUNTER after each conversion by the data acquisition board
during scanning. This signal is referred to as SCANCLK. See
Chapter 3,

Theory of Operation

, for a description of MUXCOUNTER.

Pin 43 is a reserved digital input.

The following specifications and ratings apply to the digital I/O lines.

Absolute maximum voltage

Input rating

5.5 V with respect to DIG GND

Digital input specifications (referenced to DIG GND):

V

IH

input logic high voltage

2 V minimum

V

IL

input logic low voltage

0.8 V maximum

I

I

input current leakage

±1 µA maximum

Digital output specifications (referenced to DIG GND):

V

OH

output logic high voltage

3.7 V minimum at 4 mA maximum

V

OL

output logic low voltage

0.4 V maximum at 4 mA maximum

Table 2-16. SCXIbus to SCXI-1121 Rear Signal Connector to Data Acquisition Board Pin Equivalences

SCXIbus Line

SCXI-1121

Rear Signal

Connector

MIO-16

Lab-NB/Lab-PC

Lab-PC+/Lab-LC

PC-LPM-16

MOSI

SERDATIN

ADIO0

PB4

DOUT4

D*/A

DAQD*/A

ADIO1

PB5

DOUT5

INTR*

SLOT0SEL*

ADIO2

PB6

DOUT6

SPICLK

SERCLK

EXTSTROBE*

PB7

DOUT7

MISO

SERDATOUT

BDIO0

PC1

DIN6