Chapter 2

Configuration and Installation

© National Instruments Corporation

2-11

SCXI-1121 User Manual

After selecting the excitation mode of operation desired—Voltage or
Current—as described in the previous section, use Table 2-8 to set your
SCXI-1121 for the level of operation. Your SCXI-1121 is shipped with the
Voltage mode set to 3.333 V.

Table 2-7. Maximum Load per Excitation Channel

Excitation Level

Maximum Load

3.333 V

28 mA

10 V

14 mA

0.150 mA

10 k

0.450 mA

10 k

Table 2-8. Excitation Level Jumper Selection

Excitation

Channel

Jumpers

3.333 V or 0.150 mA

(Factory Setting)

10 V or 0.450 mA

0

W16 and W26

1

W24 and W25

2

W36 and W37

3

W48 and W49

W16

1 2 3

W26

1

2

3

W16

1 2 3

W26

1

2

3

W24

1 2 3

W25

1 2 3

W24

1 2 3

W25

1 2 3

W36

1 2 3

W37

1 2 3

W36

1 2 3

W37

1 2 3

W48

1 2 3

W49

1 2 3

W48

1 2 3

W49

1 2 3