Table 2-11, Nulling resistors and corresponding channel -27 – National Instruments SCXI-1121 User Manual

Page 42

Advertising
background image

Chapter 2

Configuration and Installation

SCXI-1121 User Manual

2-26

www.natinst.com

The value of all the nulling resistors on your terminal block is 39 k

. Notice

that these resistors are socketed for easy replacement. These sockets best fit
a 1/4 W resistor lead size.

To determine your nulling range, use the following formula (refer to
Figure 2-9 for visual help):

where

R

g

is the nominal strain-gauge resistance value.

R

d

is either a completion resistor or a second strain-gauge nominal

resistance.

R

null

is the nulling resistor value.

V

exc

is the excitation voltage (3.333 or 10 V).

For example, assuming:

V

exc

= 3.333 V

R

g

= 120

R

d

= 120

R

null

= 39 k

V

nulling

= ±2.56 mV

Table 2-11. Nulling Resistors and Corresponding Channel

Channel Number

Nulling Resistor

0

R3

1

R5

2

R7

3

R9

V

nullingrange

V

exc

2

----------

V

exc

R

d

R

null

R

g

+

(

)

R

null

R

g

R

d

R

null

R

g

+

(

)

+

-----------------------------------------------------------

±

=

Advertising