RTD Mode

Excitation current ................................... 0.15 mA ±0.04%, 0.45 mA ±0.2%

Maximum load resistance ...................... 10 k

Drift ........................................................ 40 ppm/ °C

Lead resistance effect............................. Negligible (4-wire measurement)

Resistance range..................................... 10 k

, maximum

Strain Gauge Mode

Bridge types ........................................... Quarter-, half-, and full-bridge

Bridge completion.................................. Two 4.5 k

±0.05% ratio

tolerance resistors

Excitation voltage .................................. 3.333 V ±0.04% or 10 V ±0.2%

Resistance range..................................... 120

, minimum at 3.333 V

800

, minimum at 10 V

Half-bridge voltage ................................ V

EXT

/2 ±0.04%

Cold-Junction Sensor

1

SCXI-1320 and SCXI-1321

Accuracy ......................................... 1.0 °C over 0 to 55 °C
Output ............................................. 10 mV/°C

SCXI-1328

Accuracy

2

....................................... 0.35 ° from 15 to 35 °C

0.65 ° from 0 to 15 ° and
35 to 55 °C

Output ............................................. 1.91 to 0.58 V from 0 to 55 °C
Output at 25 °C ............................... 1.25 V

1

Located on the SCXI-1320, SCXI-1328, and SCXI-1321 terminal blocks

2

Includes the combined effects of the temperature sensor accuracy and the temperature difference between the temperature
sensor and any screw terminal.