Analog output circuitry, Figure 3-6. analog output circuitry, Analog output circuitry -15 – National Instruments SCXI-1121 User Manual

Page 77: Figure 3-6

Advertising
background image

Chapter 3

Theory of Operation

© National Instruments Corporation

3-15

SCXI-1121 User Manual

Analog Output Circuitry

Figure 3-6 shows the SCXI-1121 analog output circuitry.

Figure 3-6. Analog Output Circuitry

The SCXI-1121 output circuitry consists of a buffered-output multiplexer
and channel-select hardware. The channel-select hardware consists of a
two-bit counter, MUXCOUNTER. This counter is needed when the board
is operating in the Multiplexed-Output Mode. The counter output is sent to
the multiplexer address pins to determine which of the four channels is to
be connected to MCH0. In the Single-Channel Read mode, the
MUXCOUNTER is loaded with the desired channel number. In the
Scanning mode, the counter is loaded with the first channel to be read.
During the scan, the counter is clocked by SCANCLK from the data
acquisition board, or TRIG0 from the SCXIbus, depending on the state of
the CLKSELECT bit in the Configuration Register. During scanning
operations, the MUXCOUNTER is reloaded with the channel value stored

AB0

Switch

Buffer

MCH0+

Output

Mux

Channel 0
Channel 1
Channel 2
Channel 3

MCH0–

MCH1+

From

Channel 1

From

Channel 2

MCH1–

From

Channel 3

MCH2–

MCH2+

MCH3+

MCH3–

DTEMP

MCH4+

MCH4–

MTEMP

Analog

Reference

SCXIbus

Rear Signal Connector

Output

Stage

Control

Output Stage and

Hardware Scan Control