National Instruments SCXI-1121 User Manual

Page 110

Advertising
background image

Appendix A

Specifications

SCXI-1121 User Manual

A-2

www.natinst.com

Filtering (jumper-selectable) ..................4 Hz (–10 dB) or 10 kHz (–3 dB),

3-pole RC

Noise (400 kHz bandwidth)

1

Input (gain = 1,000)

4 Hz filter .................................100 nV

rms

10 kHz filter .............................4 µV

rms

Output (gain = 1)

4 Hz ..........................................150 µV

rms

10 kHz ......................................1 mV

rms

Output selection time (with 5 V step, all gains)

2

0.012% accuracy..............................5.2 µsec typical, 7 µsec maximum
0.006% accuracy..............................10 µsec
0.0015% accuracy............................20 µsec

Rise time

4 Hz .................................................0.12 sec
10 kHz .............................................70 µsec

Slew rate .................................................0.15 V/µsec

Operating common-mode voltage,

50 Hz or 60 Hz
Channel to channel or
Channel to earth...............................250 V

rms

2

Common-mode rejection ratio, 50 or 60 Hz

1 k

in input leads ..........................160 dB minimum at 4 Hz

bandwidth

NMR (50 or 60 Hz) .........................60 dB at 4 Hz bandwidth

Input protection (continuous) .................250 V

rms

maximum

Output protection ....................................Continuous short-to-ground

Power dissipation....................................7.5 W maximum

1

Includes the combined effects of the SCXI-1121 and the AT-MIO-16F-5

2

Module tested following the UL1244 standard to twice the working voltage +1000

V

rms