Parameters, Comments, Usage – VXI VT1422A User Manual

Page 315: Output:current:amplitude

Advertising
background image

VT1422A Command Reference 313

Chapter 7

OUTPut

Parameters

Comments

Select 488E-6 (or MAX) for measuring resistances of less than 8000

Ω. Select

30E-6 (or MIN) for resistances of 8000

Ω and above. <amplitude> may be

specified in µA (ua).

For resistance temperature measurements ([SENSe:]FUNCtion:TEMPerature)

the Current Source SCP must be set as follows:

When *CAL? is executed, the current sources are calibrated on the range

selected at that time.

When Accepted: Not while INITiated

Related Commands: *CAL?, OUTP:CURR:AMPL?

*RST Condition: MIN

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

OUTP:CURR:AMPL 488ua,(@116:123)

Set Current Source SCP at channels 16
through 23 to 488 µA

OUTP:CURR:AMPL 30E-6,(@105)

Set Current Source SCP at channel 5 to
30 µA

OUTPut:CURRent:AMPLitude?

OUTPut:CURRent:AMPLitude? (@<channel>)

returns the range setting of the

Current Source SCP channel specified by channel.

Parameters

Comments

The <channel> parameter must specify a single channel only.

If <channel> specifies an SCP which is not a Current Source, a +3007, "Invalid

signal conditioning plug-on" error is generated.

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

amplitude

numeric (float32)

MIN | 30E-6 | MAX | 488E-6

A dc

ch_list

channel list (string)

100 - 163

none

MAX (488

μA)

for RTD,85 | 92 and THER,2250

MIN (30

μA)

for THER,5000 | 10000

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

channel

channel list (string)

100 - 163

none