336 VT1422A Command Reference

Chapter 7

[SENSe]

[SENSe:]DATA:FIFO:MODE

[SENSe:]DATA:FIFO:MODE <mode>

sets the mode of operation for the

FIFO buffer.

Parameters

Comments

In BLOCk(ing) mode, if the FIFO becomes full and measurements are still

being made, the new values are discarded.

OVERwrite mode is used record the latest 65,024 values. The module must be

halted (ABORT sent) before attempting to read the FIFO. In OVERwrite Mode,
if the FIFO becomes full and measurements are still being made, new values
overwrite the oldest values.

In both modes Error 3021, "FIFO Overflow" is generated to indicate that

measurements have been lost.

When Accepted: Not while INITiated

Related Commands: SENSE:DATA:FIFO:MODE?,

SENSE:DATA:FIFO:ALL?, SENSE:DATA:FIFO:HALF?,
SENSE:DATA:FIFO:PART?, SENSE:DATA:FIFO:COUNT?

*RST Condition: SENSE:DATA:FIFO:MODE BLOCk

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

SENSE:DATA:FIFO:MODE OVERWRITE

Set FIFO to overwrite mode

DATA:FIFO:MODE BLOCK

Set FIFO to block mode

[SENSe:]DATA:FIFO:MODE?

[SENSe:]DATA:FIFO:MODE?

returns the currently set FIFO mode.

Comments

Returned Value: String value either BLOCK or OVERWRITE. The data type

is string.

Related Commands: SENSE:DATA:FIFO:MODE

Send with VXIplug&play Function: hpe1422_cmdString_Q(...)

Usage

DATA:FIFO:MODE?

Enter statement returns either BLOCK or
OVERWRITE

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

mode

discrete (string)

BLOCk | OVERwrite

none