Replacing the MSP/SpO

2

Boards

The following instructions describe how to replace the MSP/SpO

2

Boards in the

Expansion Module with a CO

2

option. Instructions for replacing SpO

2

Boards in SpO

2

module and Expansion Module without a CO

2

option installed are described in

“Replacing

the SpO2 Boards”

on page 98

1.

Open the Expansion Module (

“Expansion Module”

on page 90).

2.

Unplug the flex-cable connector J5 from the MSP Board, but do not attempt to
remove the Mainstream CO

2

connector side panel.

3.

Disconnect the speaker cable from the MSP Board and remove the speaker side
panel.

4.

Remove the three screws securing the MSP Board and SpO

2

Board to the rear panel

(see the figure above).

5.

Lift out the two circuit boards and SpO

2

patient connector side panel as a unit. (The

side panel easily slides out.)

6.

Using equal pressure on both sides of the connector, disconnect the patient
connector side panel connector from the SpO

2

Board.

7.

Disconnect the two circuit boards by gently pulling them apart.

When installing the MSP/SpO

2

Boards:

plug the two circuit boards together

connect the patient connector side panel connector to the SpO

2

Board

Caution When replacing the MSP/SpO

2

Boards, remove both boards with the

patient connector side panel still connected. Attempting to disconnect the
patient connector side panel from the SpO

2

Board before removing both circuit

boards with the side panel can damage the flex-cable between the side panel and
the SpO

2

Board.

Caution In the next step, the connector securing clips are fragile and will break
if not handled carefully.

Flexible

Connector J5

Speaker

Cable