Repair procedures

Welch Allyn Propaq CS Vital Signs Monitor

Closing the monitor

1.

Before closing the monitor casing, confirm the following:

Cable to P6 on the Analog Board goes behind and below the cable to P5, and over
the cable to P7.

Cable to P5 on the Analog Board goes over and in front of the cable to P6.

Cable to P7 on the Analog Board goes under the cable to P6.

Cable to P1 (the expansion cable) must not lay over the transformer.

Cable to P8 on the Digital Board rests to the right of the transformer and over the
ribbon cable and cable to P9.

Cable to P9 lies under the cable to P8.

All cables and hoses are not pinched.

The Front Chassis pressure transducer tube has been securely reconnected.

The rubber O-ring gasket is in place in the Rear Chassis.

Verify that the P10 locking clips are secure.

2.

Torque the five screws securing the Interface Chassis to the Front Chassis to 6.65 to
7.35 in-lb (0.75 to 0.83 N-m).

3.

Torque the six screws securing the Rear Chassis to the Interface Chassis to 8.0 ± 0.4
in-lb (0.9 ± 0.04 N-m).

Caution Route the red and white Display Power Cable down inside the Front
Chassis prior to the application of the Interface Chassis to avoid pinching the
Cable.

P5

P6

P7

P8

P1

P9

P3

P2

P11

P10

P12

P15

P1

P9

P10

P5

P6

P7

P8