Mainstream board, Board providing the spo, Measurement function. the co – Welch Allyn PROPAQ CS User Manual

Page 131: And spo, Service manual technical overview 127

Advertising
background image

Service Manual

Technical overview

127

Mainstream Board

The MSP Board is divided into isolated and non-isolated circuitry.

Non-Isolated Circuitry. The Non-Isolated circuitry on the MSP Board interfaces with the

monitor, the printer, the speaker, P6 and P7 expansion connectors, and with the
Mainstream Processor which resides on the isolated section of the MSP Board.

Isolated Circuitry. The Isolated Circuitry directly controls the hardware responsible for

the CO

2

measurement function.

CPU Circuitry. The CPU Circuitry includes the Philips 80C552 micro controller, 32K of

RAM, 64K of ROM, and standard RESET circuitry.

A UART directly interfaces with the pulse oximetry SpO

2

Board providing the SpO

2

measurement function.

The CO

2

and SpO

2

functions are isolated from all other subsystems in the monitor or

Expansion Modules, but not from each other.

co2 block

PRINTER

I/F CONN.

EXP.

#1 CONN.

EXP.

#2 CONN.

SPEAKER

I/F CIRC.

NON-ISO

POWER

IPC EXP.

CIRC.

ISO PWR

TRANS.

OPTO

COUPL.

OPTO

COUPL.

ISO PWR

SUPPLY

SPO2

I/F CIRC.

SPO2

XDUCER

AND

CONN.

MSP NON-ISOLATED CIRCUITRY

SSP NON-ISOLATED CIRCUITRY

MSP ISOLATED CIRCUITRY

CAPNOGRAPHY FUNCTIONAL BLOCK DIAGRAM

1097-13

PUMP

CONTROL

SIDE-

STREAM

STICK

PCB

CPU

A/D

CONV.

&

MULTIPLEX

CONTROL

INFRARED

SOURCE

DETECTOR

CONTROL

MOTOR

DRIVE

MOTOR

SENSE

(POSITION)

PRESSURE

SENSORS &

AMPLIFIERS

WATER

TRAP

DETECTOR

PUMP

FILTER

PUMP

(Pneumatic connections

not shown.)

NELLCOR

SPO2

BOARD

CPU

A-D

CONV.

INFRARED

SOURCE

DET.
BIAS

CIRC.

HEATER

CIRC.

MOTOR

DRIVE

POSN.

SENSE

DET.

SIGNAL

DET.

TEMP.

BARO.

CO2

SENSOR

AND

CONN.

EXTENDER

BOARD

CO2 SIGNAL

TEMPERATURE

Advertising