Welch Allyn PROPAQ CS User Manual

Page 53

Advertising
background image

Service Manual

Calibration

49

8.

Disconnect the DMM leads.

Table 12. Main power supply voltage checks

Supply voltages

Ref. lead

Pos. lead

Voltage limits (V)

Non-isolated

+15

TP39

TP3

+14.7 to +15.3

–15

TP39

TP11

–14.8 to -15.5

Isolated

+VR

TP43 or TP1

TP9

+4.91 to +5.09

+VS

TP43 or TP1

TP5

+6.58 to +7.15

–VS

TP43 or TP1

TP21

-6.65 to -7.15

+VB

TP43 or TP1

TP13

+5.40 to +6.10

+5.6

TP43 or TP1

TP33

+5.55 to +5.70

IBP shield

TP43 or TP1

P6, Pin 10

+2.47 to +2.53

C.L.P. detector

a

a.

Make sure the ECG input connector (P5) is disconnected from the Analog Board for this measurement.
Reconnect when done.

TP43 or TP1

U5, Pin 1

0.00 to +0.25

NIBP Pressure Offset

TP43 or TP1

TP31

+2.40 to +2.60

OSC baseline

TP43 or TP1

TP47

-0.95 to -1.25

+VA

TP43 or TP1

TP7

(+VR + 0.1 V) to
(+VR + 0.2 V)

-VA

TP43 or TP1

TP15

(+VR + 0.2 V) to
-(VR + 0.35 V)

+2.5

TP43 or TP1

TP17

+2.48 to +2.52