Service Manual

Functional verification

29

“Expansion module option configurations”

on page 8. The following table shows how to

identify compatible sensors for each Nellcor option:

Printer tests

The following steps check the functionality of the printer.

1.

Disconnect all cables from the monitor.

2.

Turn the monitor OFF then ON.

3.

Press SETUP > WAVE SELECT > INSERVICE.

4.

Press HOME

.

5.

Wait at least 8 seconds and then press the printer's SNAPSHOT key.

6.

Check that the printer prints an approximately eight-inch strip of the simulated patient
data. The paper should stop only after the entire waveform has been fed out of the
printer.

7.

Press the printer's START/STOP key. The printer should begin continuous printing.

8.

Press the START/STOP key to stop printing.

9.

Lay the monitor on its back.

10. Press the START/STOP key to start printing.

11. Completely open the paper door on the bottom of the Expansion Module.

12. Check that the printer stops printing, and that an equipment alarm and CHECK DOOR

message appear on the monitor screen.

13. Cut or tear the printer paper in the printer mechanism from the roll, leaving the strip of

paper in the printer mechanism.

14. Completely close the paper door.

15. Press the START/STOP key.

16. After the printer runs out of paper, check that an equipment alarm and PAPER OUT

message appear on the monitor screen.

17. Open the printer door and reload the paper into the printer.

Table 7. Sensor compatibility

Sensor Description

Connector

Compatible with

Pins

Notches

Color

MP-203

MP-405

Nell-2, MP-507

Single-ended

7

0

Varies

Yes

No

No

Differential

7

1

Varies

Yes

Yes

No

OxiMax

9

2

Purple

Yes

Yes

Yes

OxiMax-Exclusive

8

2

White

No

No

Yes

Note

Functional verification should be performed on the monitor before beginning the
printer tests. The monitor must be functioning properly to perform printer tests.