Replacing the pump – Welch Allyn PROPAQ CS User Manual

Page 89

Advertising
background image

Service Manual

Repair procedures

85

Replacing the pump

New wires and a connector are provided with the replacement pump kit.

1.

Remove the battery pack (

“Removing the single battery pack”

on page 66).

2.

Open the monitor (

“Opening the monitor”

on page 68).

3.

Remove the Recharger Board (

“Replacing the Recharger Board”

on page 83).

4.

Unplug the pump cable from the Recharger Board.

5.

Disconnect the tubing from the pump.

6.

Cut the tie wraps securing the pump to the Recharger Board.

7.

Lift the pump from the Board, and remove the old tie wraps from the support
brackets.

When installing a pump:

Slide new tie wraps into support brackets as shown in the figure below.

Set the pump onto the Recharger Board.

Secure the pump to the Recharger Board with tie wraps, aligning the fastener ends as
shown in the figure below and cutting off the excess.

Plug in the pump cable to connector P7 on the Recharger Board.

0696-41

P7

NIBP

Pump

Support

Bracket

Correct

Position of

Tie-wrap