Warranty service, Technical support services, Recommended service intervals – Welch Allyn PROPAQ CS User Manual

Page 10

Advertising
background image

6

Overview

Welch Allyn Propaq CS Vital Signs Monitor

Warranty service

DO NOT ATTEMPT TO REPAIR the monitor yourself during the warranty period. For
service and repair, contact Welch Allyn. See below.

Technical support services

Welch Allyn offers a wide range of technical support services including:

24-hour telephone support

loaner equipment

service contracts

field replaceable units (FRUs)

factory repair

For any of these services, contact Welch Allyn Technical Support:

Worldwide Internet Access

World Wide Web (WWW) URL:

http://www.monitoring.welchallyn.com/service

World Wide E-mail: [email protected]

Recommended service intervals

At the intervals listed below, check the Propaq CS monitor for normal operation.

USA 1-800-535-6663

France (+33) 1-60-09-33-66

Australia (+61) 2-9638-3000

Latin America (+1) 305-669-9591

Germany (+49) 7477-927-173

Singapore (+65) 6291-0882

European Call

Center

(+353) 469-067-790

Canada 1-800-561-8797

Japan (+81) 3-5212-7391

United Kingdom 0-207-365-6780

South

Africa

(+27) 11-777-7509

China (+86) 21-6327-9631

Interval/Condition

Perform

Located in this
manual

Every 6 months to 2 years (according to
hospital protocols).

Complete risk (leakage) current Safety Check
followed by a Functional Verification.

“Functional
verification”

If battery does not retain a charge.

Check battery pack capacity.

“Troubleshooting”

Monitor is dropped or suspected of
damage or rough handling.

Complete Safety Check followed by Functional
Verification.

“Functional
verification”

Suspected malfunction with all or part of
monitoring parameters.

Functional Verification of suspected parameter(s).

“Functional
verification”

Monitor does not pass Functional
Verification.

Calibration of monitoring parameter circuits, or
repair, followed by Safety Check and Functional
Verification.

“Functional
verification”

and

“Calibration”

WARNING If the monitor is opened for repair or calibration, a dielectric strength
test must be completed to ensure the integrity of the patient isolation barrier. The
dielectric strength test procedure is provided in

“Functional verification”

.