118

Technical overview

Welch Allyn Propaq CS Vital Signs Monitor

CUFF

PRIMARY

TRANSDUCER

SECONDARY

TRANSDUCER

DUMP

VALVE

CLOSED

SERVO

(LINEAR)

VALVE

CLOSED

BLEED
VALVE

OPEN

PUMP

OFF

ATMOSPHERIC PRESSURE

INCREASING PRESSURE (INFLATION)

DECREASING PRESSURE (DEFLATION)

PRESSURIZED (TARGET REACHED)

NIBP PNEUMATICS AT MAXIMUM TARGET PRESSURE

0696-64

TO DAP A-D

CONVERTER

TO DCP A-D

CONVERTER

CUFF

PRIMARY

TRANSDUCER

SECONDARY

TRANSDUCER

DUMP

VALVE

CLOSED

SERVO

(LINEAR)

VALVE

PARTIALLY

OPEN

BLEED
VALVE

OPEN

PUMP

OFF

ATMOSPHERIC PRESSURE

INCREASING PRESSURE (INFLATION)

DECREASING PRESSURE (DEFLATION)

PRESSURIZED (TARGET REACHED)

NIBP PNEUMATICS DURING MEASUREMENT BLEED

0696-65

TO DAP A-D

CONVERTER

TO DCP A-D

CONVERTER