Welch Allyn PROPAQ CS User Manual

Page 51

Advertising
background image

Service Manual

Calibration

47

8.

Verify that the input current from the variable dc power supply is less than 1.8 A and
that the green LED on the monitor’s right side panel is on.

Figure 2. Recharger PCB measurement and adjustment locations

9.

Connect the DMM positive lead to P12 pin 1 or 5.

10. Connect the DMM reference (negative) lead to non-isolated ground at TP2.

11. Adjust RP5 to 9.400 ± 0.010 V dc.

12. Move the DMM positive lead to TP4.

13. Adjust RP3 to 2.393 ± 0.015V dc.

14. Move the positive lead back to P12 pin 1 or 5.

15. Vary the input voltage from 12 to 28 V and check that the output voltage remains

constant. The battery load voltage must not vary more than 0.100 V dc.

16. Decrease the power supply voltage from 12 to 7 V while checking the right side panel

green LED. The LED should turn off between 7 and 12 V.

17. Disconnect the adapter cable from the monitor’s right side panel.

18. Remove the 4.9

Ω test load and replace it with the 4.65Ω test load.

19. Reconnect the adapter cable.

20. Adjust RP1 to 9.350 ± 0.010 V dc.

21. Remove the 4.65

Ω.

22. Reconnect the cable to P12.

CATHODE

D33

T3

VLV2

VLV1

VLV3

RP7

NIBP PUMP

P11

P10

P12

+BAT

TP1

P14

TP3

Q1

GND

TP2

OVR SET

TP4

PT2

RP5

RP3

PIN 1

RP1

RP5

TP2

TP4

RP3

RP1

P12