Removing the dual battery, In expansion module, Removing the single battery pack – Welch Allyn PROPAQ CS User Manual

Page 70: Removing the dual battery in expansion module

Advertising
background image

66

Repair procedures

Welch Allyn Propaq CS Vital Signs Monitor

Removing the single battery pack

The following steps describe how to remove the battery pack from the monitor:

1.

Using a flat blade screwdriver, unscrew the four screws securing the battery pack
cover as shown below.

2.

Disconnect the monitor cable from the battery pack cable.

3.

Remove the battery pack from the monitor.

4.

Store the battery in a safe place while disassembling the monitor.

Removing the dual battery in expansion module

The Expansion Module uses a dual battery pack. The following instructions describe how
to remove the dual battery.

1.

Using a Phillips screwdriver, remove the four screws securing the Expansion Module
to the monitor (see below).

Monitor cable

Battery pack

Battery pack cable

Battery pack cover

Note

Four screws secure the Expansion Module to the monitor. Four additional screws
secure the rear panel to the Expansion Module Chassis; do not remove these
additional screws.

Advertising