Functional verification

Welch Allyn Propaq CS Vital Signs Monitor

Reset Radio menu (5 GHz radio card)

For monitors containing a 5 GHz radio card, a concealed menu is provided that allows
resetting the 5 GHz card’s configuration back to its factory default configuration. It may be
necessary to return the radio card to its default settings if:

The wireless network parameters or security key were updated incorrectly

A network security setting was enabled that is not supported by the network
infrastructure

The network name was changed incorrectly

The default password was changed and forgotten

The Radio Card Configuration Reset screen can only be entered under the following
conditions:

The monitor is equipped with a working 5 GHz radio card

There is no Acuity connection active (Network status is set to OFF)

The 5 GHz radio card has finished its initialization and is ready to accept the Reset
Defaults command

No other 5 GHz Reset Radio Defaults request has been accepted (only one active
request allowed)

Resetting the radio to factory defaults (5 GHz radio card)

To reset the 5 GHz radio card to factory defaults, open the Network Status screen. To
access this screen, from the Main Menu press SETUP > MORE > MORE > SERVICE >
YES > MORE > MORE > RADIO
.

.

Once at the Network Status screen, press the unlabeled menu keys in the exact following
sequence: 4 > 1 > 3 (see the above graphic). The sequence must be completed within 5
seconds of pressing the first button.

NETWORK STATUS

T1

°C

PREVIOUS

MENU

BATTERY: 8.4 VOLTS

FIRMWARE VER:

NET NAME:

MONITOR MAC:

MONITOR IP:

1.00.00 A0023

pco|99

00:1A:FA:00:01:02

172.29.51.32

37.0

1

2

3

4

Note: The above button numbers and network variables are for reference only.