Service Manual

Repair procedures

93

Replacing Expansion Module front panel

The front panel contains the printer buttons. Like the rear panel, the front is sealed to the
Chassis with a rubber O-ring gasket. This gasket can be re-used when replacing the front
panel.

The Expansion Module front panel is attached to the Chassis with four screws. Use the
following procedure to remove the front panel:

1.

Remove the four screws (see the figure below) holding the front panel.

2.

For Expansion Module options with Sidestream CO

2

(226, 227, 228, or 229),

disconnect the cables from J1 and J3 on the SSP Board.

3.

Separate the front panel from the Chassis, disconnecting the front panel cable from
the Expansion Module button board.

A rubber O-ring gasket seals the front panel to the Chassis. Save this gasket to re-
install it on the front panel.

When installing the Expansion Module front panel:

Verify the O-ring gasket is in place and not pinched.

Torque screws to 6.65 to 7.35 in-lb (0.75 to 0.83 N-m)

Note

To replace the Sidestream CO

2

assemblies, see

“Replacing the Sidestream CO2

assemblies”

on page 100.

Advertising