Section 5. Calibration

Introduction

Adjustments to programming and readback accuracy for the power supply are made
using cover-off procedures which must be performed by a service technician.

Calibration Setup

Service

Environment

and

Precautions

Follow established antistatic procedures.

Work at a bench with adequate room and support for the unit under test and for

all equipment required.

To reduce shock hazard, use only an insulated, straight-blade screwdriver when

calibrating trim potentiometers.

NEVER calibrate these units alone. Ensure that someone is present who can help

should an accident occur.

Equipment

Required

You will need the following equipment when calibrating your power supply.

5 1/2 digit true RMS voltmeters (DVMs) (±0.05% accuracy) with test leads

Current shunt 1 mV/A (±0.25%) and connecting wire. The recommended

current ratings for the DC shunt and connecting wire must be at least 10%

greater than the power supply's output current rating. Please see Table 2.4,

“Current Carrying Capacity for Load Wiring,” on page 45.

Non-conducting, straight-bladed screwdriver

WARNING
Exercise caution when using and calibrating a power supply. High energy levels
can be stored at the output voltage terminals on a power supply in normal
operation. In addition, potentially lethal voltages exist in the power circuit and on
the output and sense connectors of a power supply with a rated output greater
than 40 V. Filter capacitors store potentially dangerous energy for some time
after power is removed.