Section 5. Calibration

Introduction

Adjustments to programming and readback accuracy for the power supply are made
using cover-off procedures which must be performed by a service technician.

Calibration Setup

Service

Environment

and

Precautions

•

Follow established antistatic procedures.

•

Work at a bench with adequate room and support for the unit under test and for

all equipment required.

•

To reduce shock hazard, use only an insulated, straight-blade screwdriver when

calibrating trim potentiometers.

•

NEVER calibrate these units alone. Ensure that someone is present who can help

should an accident occur.

Equipment

Required

You will need the following equipment when calibrating your power supply.

•

5 1/2 digit true RMS voltmeters (DVMs) (±0.05% accuracy) with test leads

•

Current shunt 1 mV/A (±0.25%) and connecting wire. The recommended

current ratings for the DC shunt and connecting wire must be at least 10%

greater than the power supply's output current rating. Please see Table 2.4,

“Current Carrying Capacity for Load Wiring,” on page 45.

•

Non-conducting, straight-bladed screwdriver

WARNING
Exercise caution when using and calibrating a power supply. High energy levels
can be stored at the output voltage terminals on a power supply in normal
operation. In addition, potentially lethal voltages exist in the power circuit and on
the output and sense connectors of a power supply with a rated output greater
than 40 V. Filter capacitors store potentially dangerous energy for some time
after power is removed.