3 calibration, Calibration – American Magnetics 185 & 186 Liquid Level Instruments (CE-Marked) User Manual

Page 27

Advertising
background image

Rev. 3

17

3 Calibration

Model 185/186 instruments are calibrated at the factory for a specific

length sensor for use in a specific liquid. The calibration length and

calibration liquid are listed on the calibration sticker on the bottom of the

instrument. If the factory calibration method utilized was approximate,

the calibration length will be noted as an approximate value.

3.1 Relationship between Calibration and Sensor Length

The capacitance-based method of measuring the liquid level operates by

measuring the frequency of an oscillator, which is contained in the

oscillator/transmitter unit. As the liquid level varies, the value of the

capacitance varies proportionally. Since the dielectric properties of liquids

vary and the component tolerances for the sensor and oscillator introduce

variations, a calibration is required to assure maximum accuracy for a

specific sensor immersed in the target liquid. The calibration MIN and

MAX settings correspond to the maximum and minimum oscillation

frequencies, respectively, for a given sensor and target liquid

configuration.

The LENGTH setting of the instrument is only provided as a means of

scaling the 0% (MIN) to 100% (MAX) range of the measurement to

meaningful units of length. During the calibration it is important to

accurately measure the distance between the physical locations on the

sensor corresponding to the MAX and MIN calibration points. The

measured value for the length will be used in configuring the instrument

for operation.

3.2 Variations in the Dielectric with Changing Density

For cryogenic liquids, the dielectric of the liquid will change with a change

in density. The amount of change is dependent on the properties of the

specific liquid. Figure 3-1 illustrates the variations in dielectric for

nitrogen vs. pressure under saturated conditions.

1

Since the instrument

uses a capacitance-based method for determining liquid level, such a

change in the dielectric of the liquid will result in a shift in the level

reading of the instrument. The calibration procedures described herein are

most accurate when applied in situations where the operating conditions

of the cryo-vessel are relatively constant, i.e. the operating pressure and

temperature of the cryo-vessel are relatively constant.

1. Data obtained from NIST Standard Reference Database 12.

Advertising