Remote Interface Reference

IEEE-488 Communication DIP Switch Settings

5.2.3

IEEE-488 Communication DIP Switch Settings

The 5 DIP switches located on the rear panel of the Model 185/186 are

used to set the primary IEEE-488 bus address of the unit.

5.2.3.1 IEEE-488 primary bus address

The Model 185/186 primary bus address is controlled by switches 1

through 5 of the communication DIP switch on the rear panel. Valid

primary addresses are between 0 and 30. The Model 185/186 does not use

secondary addressing. Note that many IEEE-488 controller cards in

external computers will use address 0 (or 21). The bus address for each

Model 185/186 should be unique with respect to other Model 185/186 units

or any other devices on the bus. The switch settings for the various

addresses are (on = 1 or the up position):

DIP switch

Primary bus

address

1

2

3

4

5

off

off

off

off

off

0

off

off

off

off

on

1

off

off

off

on

off

2

off

off

off

on

on

3

off

off

on

off

off

4

off

off

on

off

on

5

off

off

on

on

off

6

off

off

on

on

on

7

off

on

off

off

off

8

off

on

off

off

on

9

off

on

off

on

off

10

off

on

off

on

on

11

off

on

on

off

off

12

off

on

on

off

on

13

off

on

on

on

off

14

off

on

on

on

on

15

on

off

off

off

off

16

on

off

off

off

on

17

on

off

off

on

off

18

1

0

1 2 3 4 5