2 interacting with the running vi, Virtual instrument operation – American Magnetics 185 & 186 Liquid Level Instruments (CE-Marked) User Manual

Page 64

Advertising
background image

54

Rev. 3

Virtual Instrument Operation

RS-232 Virtual Instrument

6.1.2

Interacting with the running VI

While the VI is running the user may manipulate the virtual toggle and

rotary switches in the same manner as required for the front panel

operation of the actual instrument. See the Operation section of this

manual for instructions on operating the front panel controls, however,

please note that there are some minor differences discussed below.

The RAISE/LOWER toggle switch functions slightly different in the VI. If

the RAISE/LOWER toggle switch is moved from the center position to the

RAISE or LOWER position, then the display changes to show the

appropriate parameter. After approximately 4 seconds in the RAISE or

LOWER position, the display will begin incrementing or decrementing by

tenths. After approximately 12 additional seconds, the display will begin

incrementing/decrementing by ones. Move the RAISE/LOWER toggle

switch back to the center position to stop the incrementing or

decrementing function.

The virtual instrument’s FILL LED indicator is only accurate if the fill

toggle switch is in the AUTO position. There is no remote monitoring or

control of the manual override states of the fill toggle switch available

through the communication command set.

As a more convenient option for controlling the settings, the user may

scroll to the area above the VI and enter the values for the Sensor Length,

Hi Setpoint, A Setpoint, B Setpoint, Lo Setpoint, and Fill Interval directly

in the control fields (please observe the specified units). Any changes in

the fields are recognized and sent to the actual instrument in the form of

the appropriate command string. Any settings changed by the VI virtual

panel toggle switches or control fields are saved in permanent memory in

the actual instrument.

The VI may be gracefully stopped by using the STOP toggle switch in the

lower left corner of the VI. After stopping the VI, this switch must be

placed back in the up position in order to restart the VI.