4 Operation

The Model 185/186 and sensor (if applicable) were functionally tested and

calibrated at the factory. The calibration sticker located on the bottom of

the instrument shows the calibration length, calibration liquid, and

whether an approximate calibration method was utilized at the factory. In

the event that the calibration is incorrect for the application, the

instrument will need to be recalibrated by the user with a specific sensor

and liquid. Refer to the Calibration section for the specific procedures.

4.1 Normal Operations

4.1.1

Energize the instrument

After completion of the Installation procedures, energize the instrument

by placing the power toggle switch in the POWER position. The LED

display will briefly display AAAA and then indicate the liquid level, and the

yellow ACTIVITY LED will begin blinking.

Note

The ACTIVITY LED provides visual indication that the
microprocessor is making sensor readings. If a fault should develop
which prohibits the microprocessor from operating correctly (such as
a break in cabling) the LED will not blink or blink slowly, and the
display will continuously show 100%.

Note

If the displayed level reading is below the LO SETPOINT level or
exceeds the HI SETPOINT, an audible alarm will sound. To silence
the alarm, rotate the control mode rotary switch on the front panel to
the SILENCE position.

The instrument is normally calibrated at the factory for the specific sensor

supplied with the unit for use in a target liquid. If the need arises for

recalibration, see the Calibration section.

4.1.2

Configure the active length setting

After calibration, the instrument must be configured for the active length

of the sensor in order to scale the measurement to meaningful units of

length for display. For a standard calibration, the value of the active

length is the sensor length between the bottom hole to 1 inch below the top

hole of the sensor assembly. If the user performed a calibration, then the