3 configure the hi setpoint and the lo setpoint, 3 configure the hi setpoint and the, Lo setpoint – American Magnetics 185 & 186 Liquid Level Instruments (CE-Marked) User Manual

Page 40: Operation

Advertising
background image

30

Rev. 3

Operation

HI/LO setpoints

physical distance between the locations of the MIN and MAX calibration

points on the sensor is the active length.

The instrument allows the user to display the liquid level in units of length

(inches or centimeters) in addition to a percentage. The instrument was

shipped with the length value set to the active sensor length if a sensor

was purchased with the instrument.

To view the present length setting, place the units mode toggle switch in

either the INCH or CM position. Place the control mode rotary switch on

the front panel to the LENGTH position. Push and release the RAISE/

LOWER toggle switch either up or down. The display will momentarily

show the current length setting.

To change the length setting, use the RAISE/LOWER toggle switch to

move the setting up or down by continuously holding it in the up or down

position. The display will move slowly at first and then faster. Once near

the desired value, simply release the switch momentarily and then resume

changing the setpoint at the slower speed. The new active sensor length is

permanently stored in memory. Check the value by momentarily placing

the toggle switch in either position from the center position.

Note

The LENGTH adjustment can only be performed in the INCH or
CM units modes. The LENGTH adjustment is inactive if the units
are set for %.

4.1.3

Configure the HI SETPOINT and the LO SETPOINT

To adjust the HI and LO setpoints, place the control mode rotary switch in

the HI SETPOINT position or the LO SETPOINT position, respectively.

Use the RAISE/LOWER toggle switch to adjust the respective setpoint in

the same manner as described for the LENGTH adjustment in step 2. The

setpoints may be located anywhere between 0% to 100% of the active

sensor length. The HI and LO setpoint adjustments are compatible with

all three units modes.

a. When the measured liquid level exceeds the HI setpoint, the HI

LEVEL LED on the front panel is energized and a set of relay

contacts are closed on the 9-pin D connector J2 on the rear

panel (see the Appendix for the pinout). When the level reaches

or falls below the HI setpoint, the LED is extinguished and the

relay contacts open.

b. When the measured liquid level falls below the LO setpoint, the

LO LEVEL LED on the front panel is energized and a set of

relay contacts are closed on the 9-pin D connector J2 on the rear

panel (see the Appendix for the pinout). When the level reaches

Advertising