3 error codes, Remote interface reference – American Magnetics 185 & 186 Liquid Level Instruments (CE-Marked) User Manual

Page 60

Advertising
background image

50

Rev. 3

Remote Interface Reference

Error Codes

5.3 Error Codes

The Model 185/186 returns specific error codes for invalid commands and/

or arguments. If an error condition is returned, the command is not

processed and the configuration of the instrument is not modified. The

table below provides a list of error codes, their meaning, and any

associated limits.

Error Code

Meaning

Valid Range

1

LO setpoint out of range

0

LO

LENGTH

2

B setpoint out of range

0

B

<

A

3

A setpoint out of range

B < A

LENGTH

4

HI setpoint out of range

0

HI

LENGTH

5

Attempted to set or query for

LENGTH in PERCENT units mode

6

Invalid argument,

value

out of

maximum calibration range

1 cm

value

650 cm

7

INTERVAL setting out of range

0

INTERVAL

600 min

8

Unrecognized command

9

Invalid argument,

value

was negative

or non-numeric

0

Approximate calibration

factor

out of

range

0.1

factor

999.9