Calibration

Approximate calibration

Note

The last approximate calibration factor is not retained in the
instrument memory, therefore the effects of repeated approximate
calibrations are cumulative.

Example: Purchased a 100" active length sensor for operation in
liquid argon at atmospheric pressure, however only liquid nitrogen
is available for calibration at a maximum depth of 30":

First, the sensor is dipped as far as possible into the liquid nitrogen

and cooled. The MIN point is then set as outlined in step 2. The

MAX point is set as outlined in step 3 while the sensor is

submerged 30" in liquid nitrogen. The dielectric constant for liquid

nitrogen is 1.454 and for liquid argon is 1.53. Substituting all

values into the approximate calibration factor equation yields:

A value of 389.1 would be entered as the approximate calibration

factor as outlined in steps 6 and 7. The sensor is now approximately

calibrated for 100" active length operation in liquid argon.

8. The sensor can now be installed in the dewar containing the target

liquid. The approximate calibration can be used until an open

dewar or closed dewar calibration can be performed with the target

liquid.

9. Proceed to the Operation section for directions for configuring the

instrument.

Approximate Calibration Factor

1.53

1

1.454

1

----------------------

100

Ч

100

30

---------

389.1

=

=